Emerson In-Exsufflator Specifications

James Lubin (jlubin(AT)wport.com)
Sun, 23 Feb 1997 17:30:46 -0800

This is a multi-part message in MIME format.

--------------7AA5AA57D7F
Content-Type: text/plain; charset=us-ascii
Content-Transfer-Encoding: 7bit

Thought some of you may be interested in this device.

http://www.jhemerson.com/inexspec.htm

--------------7AA5AA57D7F
Content-Type: text/plain; charset=us-ascii; name="inexspec.htm"
Content-Transfer-Encoding: 7bit
Content-Disposition: inline; filename="inexspec.htm"

Emerson In-Exsufflator

Description:

The In-Exsufflator (Cough Machine) assists patients in clearing retained
broncho-pulmonary secretions by gradually applying a positive pressure to
the airway, then rapidly shifting to a negative pressure. This rapid shift
in pressure, via a face mask or mouthpiece produces a high expiratory flow
rate from the lungs, simulating a cough. This technique is referred to as
"Exsufflation With Negative Pressure", (E.W.N.P.) This non-invasive
technique is clinically proven to be effective.

Those who might benefit from the use of the In-Exsufflator include any
patient with an ineffective ability to cough due to poliomyelitis, muscular
dystrophy, myasthenia gravis or other neurologic disorders with some
paralysis of the respiratory muscles, such as spinal cord injury. It may
also be used to treat ineffective cough due to other bronchopulmonary
diseases, such as emphysema, cystic fibrosis, bronchial asthma and
bronchiectasis.

Use of the In-Exsufflator is particularly important for patients on
non-invasive ventilation, since suctioning is difficult without a trach
tube. Utilizing E.W.N.P. enhances or replaces the patient's natural removal
of secretions.

The In-Exsufflator comes in two models: The 2-CA which can automatically
cycle from positive to negative pressure or can be manually cycled and the
2-CMH which is only cycled manually.

Specifications:

* Maximum positive pressure: 44 mmHg (60 cm of water)
* Maximum negative pressure: 44 mmHg (60 cm of water)
* Maximum inspiratory flow: 3 liters/sec.*
* Maximum expiratory flow capacity: 10 liter/sec.**
* Mode of operation: Model 2-CA - automatic and manual timing; Model
2-CMH - manual timing
* Inspiratory, expiratory and pause times: Model 2-CA - 0 to 5 seconds;
Model 2-CMH - user variable
* Filter: Vital-signs bacterial/viral filter (5099) or equivalent
* Mask: Vital-signs face mask (5252) or equivalent
* Blower type: Two stage centrifugal blower with AC/DC universal motor
* Input power: 110 volts, 60Hz, 300 watts
* Dimensions: Model 2-CA - 11"H x 10"W x 12"D; Model 2-CMH - 11"H x 8"W x
10"D
* Weight: Model 2-CA - 22 lbs.; Model 2-CMH - 16 lbs.

* When used with inhalation restrictor. Without restrictor, inhalation flow
is the same as exhalation flow.
** Actual flow depends on maximum pressure and on airway resistance